Cooper-instruments-systems Model 3i Basic Force/Torque Indicator Manuel d'utilisateur Page 16

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 24
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 15
Model 3i Digital Force/Torque Indicator User’s Guide
15
If failed:
5. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.
6. The display will appear as follows:
Optionally exercise the sensor several times (at full scale, if possible), then press ENTER.
7. The display will appear as follows:
Apply a weight equal to the full scale of the instrument, then press ENTER.
8. After displaying “PLEASE WAIT…” the display will appear as follows:
CALIBRATION
COMPRESSION
GAIN ADJUST
APPLY FULL SCALE LOAD
10.000 LBF +/-20%
THEN PRESS ENTER
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
COMPRESSION
Optionally exercise
load cell a few times.
THEN PRESS ENTER
CALIBRATION
OFFSET
Sen.Offset Adj.Failed
Ana.Offset Adj.Failed
CALIBRATION
OFFSET
Sen.Offset Adj.Passed
Ana.Offset Adj.Passed
CALIBRATION
OFFSET
Please wait…
Vue de la page 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Commentaires sur ces manuels

Pas de commentaire