14.0 READING CONFIGURATION MENU:
RdG.CNF ENTER READING CONFIGURATION MENU:
Press ‘MENU 1) Twice though Main Menu items and display shows "RdG.CNF ".
Press ‘RESET/ENTER 2) Display shows "Rd.SC.OF " Reading Scale & Offset menu.
Press ‘MENU’ Configuration 3) Scroll through the following sequence of available Reading Menu selection list:
"Rd.SC.OF" (Reading Scale & Offset)
"Unt.tMP" (Temperature Unit)
"dEC Pt" (Decimal Point)
"FILtER" (Reading Filters)
"FLS.dSP" (Flashing Display)
"bRIGHt" (Display Brightness Option).
14.1 RD.SC.OF Enter READING SCALE & OFFSET Submenu:
Press ‘RESET/ENTER 4) To enter Reading Scale & Offset configuration Submenu if display shows "Rd.SC.OF"
(First Item of "RdG.CNF" Menu).
Press ‘RESET/ENTER 5) Display shows and flashes "dISAbL" disable (as default, Reading Scale=1.0 and
Reading Offset=0.0).
Press ‘/MAX 6) Display flashes either "ENAbLE" or "dISAbL". Select your choice and press
RESET/ENTER’. If "dISAbL" selected, display advances to "UNt.tMP" Temperature Unit
selection Submenu.
* If you select ENABLE, the following Steps 7 through 32 describe Reading Scale & Offset Format Submenu.
Otherwise continue with "UNt.tMP" in next section.
* For more detail, example, formula and Calculation Table, Refer to Section 18 of MULTI-POINT SCALE &
CF 156 36 M2544/N/0505