Series 7 Digital Force Gauges User’s Guide
30
If failed:
6. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press ENTER.
7. The display appears as follows:
Optionally exercise the load cell shaft several times (at full scale, if possible), then press ENTER.
8. The display appears as follows:
Apply a weight equal to the full scale of the instrument, then press ENTER.
9. After displaying “Please wait…” the display appears as follows:
Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.
CALIBRATION
COMPRESSION
Gain adjust
Apply full scale load
10.000 lbF +/-20%,
then press ENTER.
CALIBRATION
COMPRESSION
Ensure no load,
then press ZERO.
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures,
then press ENTER.
CALIBRATION
COMPRESSION
Optionally exercise
sensor, then press
ENTER.
CALIBRATION
OFFSET
Sensor failed
Analog failed
CALIBRATION
OFFSET
Sensor passed
Analog passed